2 Preparing and operating the CM-30S
34
October 7, 2014, 715004178 Rev. B
5. Plug the leak sensor connector into the receptacle on the front of the
module.
6. Power-on the CM-30S.
7. In the instrument, select System from the system tree.
8. Click Control > Leak sensors.
9. Enable the CM-30S leak sensor.
Installing the waste tubing
See also:
For additional information on configuring waste tubing, see the
Waters ACQUITY Advanced Polymer Chromatography (APC) System Site
Preparation Guide
or the
ACQUITY UPC
2
Site Preparation Guide
.
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
Waste tubing
Warning:
To avoid personal contamination with biologically
hazardous or toxic materials, wear clean, chemical-resistant,
powder-free gloves when installing or removing the waste tubing.
To avoid spreading contamination, do not allow contaminated
gloves to come in contact with uncontaminated surfaces, and
discard gloves immediately upon completing the procedure.
Leak sensor
connector receptacle