2 Preparing and operating the CM-30S
32
October 7, 2014, 715004178 Rev. B
Installing the leak sensor
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
Leak sensor
To install the leak sensor:
1. Power-off the CM-30S.
Warning:
To avoid the harmful effects of personal contact with solvents,
including inhalation, observe Good Laboratory Practice when you
handle them. See the Material Safety Data Sheets for the solvents you
use.
Warning:
To avoid personal contamination with biohazards or
compounds that are toxic, wear clean, chemical-resistant,
powder-free gloves when performing this procedure.
Notice:
To avoid damaging the electronic components or circuitry
of a system module, do not disconnect an electrical assembly from
the module while the module remains connected to the ac supply
source.
Follow this procedure to completely interrupt power to the module:
1. Set the module’s power switch to Off.
2. Disconnect the module’s electrical supply cord from the ac
source.
Afterward, wait 10 seconds before disconnecting any assembly
from the module.