Maintaining the Autosampler
Replacing a syringe
The instrument is fitted with a 500-µL syringe as standard equipment, but a
2500-µL syringe can be installed as part of the prep kit.
To replace a syringe
1.
From the Maintain menu of the 2707 Console, click Replace > Syringe.
Result:
The syringe plunger lowers.
2.
Unscrew the syringe from syringe valve.
Requirement:
Keep the connector in the valve intact.
3.
Disconnect the plunger from the syringe drive by sliding it straight
backward, out of the holding clip.
Syringe valve
Holding clip