Making Ethernet connections
2-11
2.
Place the wash solvent bottle to the left of the instrument.
Tip:
The approved wash solvent bottle is made of glass.
3.
Place the end of the wash solvent tubing in the filled wash solvent bottle.
Recommendations:
•
To achieve the necessary length, replace the wash solvent line with
tubing from the Waters inline degasser. Attach the tubing to the
front of the syringe valve, and route it out one of the square cutouts
on the instrument’s side.
•
To avoid flow problems, use tubing with low gas permeability and
an internal diameter sufficient to minimize flow impedance, such as
the tubing supplied with the Waters inline degasser.
4.
From either the Control menu in the 2707 Console or as a right-click
command from the 2707 Control panel in Run Samples of Empower,
select Wash/Prime Needle to fill the tubes with liquid.
Tip:
To remove all air from the syringe and achieve maximal operation,
perform an extra wash.
Tip:
If bubbles form in the syringe on the intake stroke of the wash cycle,
the wash solvent can be insufficiently degassed or, in the case of an
external degasser, the inlet tubing diameter can be too small.
Making Ethernet connections
1.
Connect one end of one Ethernet cable to the network switch, and then
connect the other end to the Ethernet card on the computer.
Recommendation:
Use only the shielded Ethernet cables in the startup
kit.
2.
Connect one end of one Ethernet cable to the instrument’s rear panel,
and then connect the other end to the network switch.
Recommendation:
For workstations with only one Ethernet instrument,
connect an Ethernet cable to the PC’s network interface card (NIC).
Requirement:
Older, nonautosensing NICs require shielded cross-over
cables, which are not supplied in the startup kit.