June 11, 2015, 715004752 Rev. A
Page 61
3.2.4.1
Operating the Trace and Scale functions
Using the Trace function, you can display an absorbance trace for the last
n
minutes
(as many as 60) of detector operation.
•
When you press the TRACE key, the detector displays the absorbance acquired
over the last 30 minutes by default. It updates the trace once every 20 seconds.
•
When you press the Scale key (Shift TRACE), the detector displays the scaled
trace with T1 (ending time) displayed (–30 for the last 30 minutes) by default.
Table 3–2:
Primary and secondary function (method) parameters
Function
Screen
Type
Units
Range
Default
λ
(Wavelength)
1
(Absorbanc
e screen)
Numeric
nm
Integer 190 nm to
700 nm
254 nm
AUFS
1
Numeric
AUFS
0.0001 to 4.0000
2.0000
Analog rate
2 (of 4) or
2 (of 5)
Choice
Hz
(
λ
): 10, 20, 40,
80
(
λλ
): 1 or 2
10
1
Filter time
constant
2 (of 4) or
2 (of 5)
Numeric
Sec
Slow
Normal
Fast
Other:
(
λ
): 0.0125 to 5.0
(
λλ
): 0.5 to 5.0
Off to disable
filtering
Normal
(for
Other:)
1.000
Data out
(single
λ
)
3 (of 4)
Choice
None
Absorbance A
Absorbance
A
Data out (dual
λ
)
3 (of 5)
Choice
None
Absorbance A
Absorbance B
Maxplot A, B
Difference A-B
Ratio A/B
Absorbance
A
Autozero on
inject
3 (of 4) or
3 (of 5)
Check
box
None
Selected/
Not selected
Checked
Autozero on
λ
changes
3 (of 4) or 3
(of 5)
Choice
None
To baseline
To zero
Disable
To zero
Voltage offset
4 (of 4) or
4 (of 5)
Numeric
mV
Integer –2000 to
2000
0
Chart polarity
4 (of 4) or
4 (of 5)
Choice
None
+
–
+
Minimum AU
5 (of 5)
Numeric
AU
0.0001 to 4.0000
0.1000
Minimum ratio 5 (of 5)
Numeric
None
0.00 to 999.99
0.00
Maximum
ratio
5 (of 5)
Numeric
None
0.00 to 999.99
2.00
Содержание 2489
Страница 2: ...June 11 2015 715004752 Rev A Page ii...
Страница 8: ...June 11 2015 715004752 Rev A Page viii...
Страница 14: ...June 11 2015 715004752 Rev A Page xiv...
Страница 46: ...June 11 2015 715004752 Rev A Page 46...
Страница 124: ...June 11 2015 715004752 Rev A Page 124...
Страница 135: ...June 11 2015 715004752 Rev A Page 135 Lower limit of temperature Temperature limitation Symbol Description...
Страница 136: ...June 11 2015 715004752 Rev A Page 136...