Installing the Detector
Site selection and power requirements
Site selection
Install the detector in an area that meets the requirements listed in the table
below.
Requirement:
You must mount the detector on a level surface to allow proper
function of the drip management system (drain tube), which you can connect
to a waste reservoir to divert solvent leaks from the flow cell.
Installation site requirements
Parameter
Requirement
Operating temperature
range
4 to 40
°
C (39 to 104
°
F)
Storage temperature
range
–30 to 60
°
C
Relative humidity
20% to <95%, noncondensing
Storage humidity range
0 to <95%, noncondensing
Bench space
12.7 cm (5 inches) clearance at rear
Vibration
Negligible
Static electricity
Negligible
Power
Grounded ac, 100/240 Vac, 50/60 Hz
Power cord type required:
• SVT in the United States
• HAR type (or better) in Europe
For information regarding the type of
cord to use in other countries, contact
your local Waters distributor.