3-22
Operating the Detector
Operating the trace and scale functions
The trace function allows you to display an LSU signal for the last
n
minutes
(up to 60) of detector operation.
•
Press TRACE to display the LSU signal acquired over the last 30
minutes by default. The LSU signal updates once every 20 seconds.
•
Press Scale (press Shift, TRACE) to display the scaled trace with T1
(ending time) displayed (-30 for the last 30 minutes) by default.
Change the ending time parameter to any number from 3 to 60. You can
use the Scale function to zoom in on a particular section of the trace.
To display the Scale parameters
1. Press Scale.
2. Press Next to display T2 (starting time). The default is 0.
3. Press Next again to display LSU1 (starting or low light-scattering
signal). The default is auto-scaled.
4. Press Next again to display LSU2 (ending or high light-scattering
signal). The default is auto-scaled.
By entering appropriate times and light scattering numbers in the four
scaling parameter boxes, you can zoom in on one section of the current
light scattering trace.
•
For LSU1 and LSU2, press CE
to reset to auto-scaled.
•
T1 represents the left-hand side of the trace, or ending time, to be
displayed. The default is -30.
•
T2 represents the right side of the trace, or starting time. The
default is 0.
Содержание 2424
Страница 8: ...viii...
Страница 30: ...1 16 2424 ELS Detector Optics Principles...
Страница 64: ...2 34 Setting up the Detector...
Страница 110: ...3 46 Operating the Detector...
Страница 126: ...4 16 Maintaining the Detector...
Страница 182: ...A 14 Safety Advisories...
Страница 186: ...B 4 Specifications...
Страница 196: ...Index 10...