Setting up a run
3-17
Setting up a run
Start a run by pressing the Run/Stop key. Before starting a run, you must set
the detector operating parameters before data can be acquired.
To set up a run
1. Press HOME
to return to the Home screen.
2. Set the drift tube heater temperature (see
3. Set the gain and gas pressure (see
).
4. Equilibrate the detector for about an hour.
5. Make sure the stop flow output switch is reset (see
You need to program several other parameters, depending on the functions
you want to perform during a run. See
fields, screen number, type of function, display units, allowable ranges, and
default settings for the Home screen and the secondary function screens. You
can view the properties of common chromatographic solvents in the table on
.
Threshold
switch mode
On, Off, Pulse,
Rectangular Wave
On
PULSE
switch mode
period
sec
0.1 to 60.0 sec
0.1 sec
RECT
WAVE
switch mode
period
sec
0.1 to 60.0 sec
0.2 sec
Number of
timed
events
0 to 50
0
Primary and secondary function (method) parameters (Continued)
Function
Units
Range
Default