3-4
Operating the Detector
If confidence tests fail
If one or more of the confidence tests fail during initial startup, the detector
behaves as follows:
•
Beeps
•
Displays an error message
•
For serious errors, displays the word “Error” in brackets (<Error>) in
place of the runtime light scattering units on the Home screen.
“Startup error messages” on page 5-2,
for a list of startup confidence tests
failures, error messages, and recommended recovery actions. Error messages
that can appear as you operate the detector and suggested corrective actions
are described in the table on
for hardware-related
causes of confidence tests failure during startup and their corrective actions.
Using the display
The detector employs a 128 × 64 bitmap graphic display and a 24-key
membrane keypad for the operator interface. After the startup confidence
tests have run successfully, the detector displays the Home screen.
Detector Home screen
You can recall the Home screen at any time by pressing HOME. At the first
use of the detector, the Home screen shows the default settings for the gain
and pressure.
Nebulizer
temperature
display
Gain
Error or additional
function modes
Light scattering
units
Lamp on/off
Shift on/off
Keypad lock/unlock
Local method
number/remote control
Next screen
Run time (minutes)
Gas pressure
Drift tube
temperature
display
Sticky diagnostics
on/off
Содержание 2424
Страница 8: ...viii...
Страница 30: ...1 16 2424 ELS Detector Optics Principles...
Страница 64: ...2 34 Setting up the Detector...
Страница 110: ...3 46 Operating the Detector...
Страница 126: ...4 16 Maintaining the Detector...
Страница 182: ...A 14 Safety Advisories...
Страница 186: ...B 4 Specifications...
Страница 196: ...Index 10...