Making signal connections
2-23
Network installation guidelines
Configurations for multiple Waters instruments use a dedicated LAN, which
requires a design based on the following guidelines:
•
100 base-T, 100-Mbps shielded twisted-pair (STP) cable
•
A maximum distance of 100 meters (328 feet)
Requirement:
You must use a network switch for multiple Ethernet
instruments. Use of a network hub in place of a network switch is not
supported.
Multiple Waters Ethernet instrument connections
Connecting to a Waters MassLynx system using Ethernet
When connecting to a MassLynx system using Ethernet, the various
components should be connected as shown in the figure below.
TP02075
Network switch
Mass
spectrometer
ELS detector
Analog input
Workstation
100 Base-T
Ethernet cable
Instrument
network card
eSAT/IN module
Analog input