Graphical Digital Panel Meter WPMZ-1 INSTRUCTION MANUAL
48
IM-0882-03
NormalHold: As long as the external control terminal is set to the L level, the display continues to be
updated with the peak to peak value. By setting the pin to H level, holding is canceled
and the latest display value is displayed.
AreaHold: As long as the external control terminal is set to the L level, the display continues to be
updated with the peak to peak value. When the terminal is set to H level, it keeps holding
at the peak to peak value at that point.
* When the display value becomes OVER during control of the peak to peak value holding function, the OVER
display will not disappear unless control is canceled once.
DevHold
This function holds value that is the most distant from reference value. The reference value is set with
“DevBaseValue”.
It is used to indicate the position of the error with respect to the target control value.
*Note that this function holds not difference value but display value.
There are two modes of “NormalHold” and “AreaHold”, and the mode can be changed by the “HoldMode”.
NormalHold: As long as the external control terminal is set to the L level, the display continues to be updated
with the most distant value. By setting the pin to H level, holding is canceled and the latest display
value is displayed.
AreaHold: As long as the external control terminal is set to the L level, the display continues to be
updated with the most distant value. When the terminal is set to H level, it keeps holding
at the most distant value at that point.
AveHold
This function performs additional moving average while running.
The average number of times is set by the “AveHoldCount” of the setting item.
It is effective in temporarily canceling vibration influences etc. by the external environment.
There are two modes of “NormalHold” and “AreaHold”, and the mode can be changed by the “HoldMode”.
NormalHold: As long as the external control terminal is set to the L level, the display continues to be
updated with the moving average value. By setting the pin to H level, holding is canceled
and the latest display value is displayed.
AreaHold: As long as the external control terminal is set to the L level, the display continues to be
updated with the moving average value. When the terminal is set to H level, it keeps holding
at the moving average value at that point.
Time
Display value
Sensor value
On the section
Time
Display value
Sensor value
On the section
Reference value
Hold the most distant