PGT
®
120.COM
Personnel Grounding Tester
Wrist Strap and Footwear Tester with serial port
User's Manual
Страница 1: ...PGT 120 COM Personnel Grounding Tester Wrist Strap and Footwear Tester with serial port User s Manual...
Страница 2: ...6 4 3 Schuhpr fung einzeln 7 4 4 Gelenkband und Schuhpr fung gemeinsam 7 4 5 Schuhpr fung als Reihenschaltung 8 5 KONFIGURATION 9 6 DATENAUSGABE BER RS232 10 7 ANSCHL SSE 12 8 MONTAGEANLEITUNG WANDKO...
Страница 3: ...100 PGT120 COM 3 17 2016 09 07 4 4 Wrist strap and footwear test 7 4 5 Footwear in series 8 5 CONFIGURATION 9 6 DATA OUTPUT VIA RS232 10 7 CONNECTORS 12 8 WALL MOUNTING INSTRUCTIONS 14 9 SPECIFICATION...
Страница 4: ...e measuring circuits The order of tests is random Footwear measurement can be configured to measure in series with hands free for passenger gates Visual and audible test results serial port and a dry...
Страница 5: ...he unit to a wall The power is supplied by a power supply Use only an original power supply connected to the AC12V socket on the rear Do not connect any conducting articles with PGT 120 COM exept orig...
Страница 6: ...Fail Red LED flashes audible signal Above the resistance upper limit Lo Fail Red LED flashes audible signal Below the resistance lower limit not applicable if lower limit is disabled Release the elect...
Страница 7: ...wer limit is disabled Lo Fail left Red LED flashes audible signal Left shoe below the resistance lower limit not applicable if lower limit is disabled Release the electrode 4 4 Wrist strap and footwea...
Страница 8: ...ss the Shoe electrode on the instrument to identify it Stand with both feet onto the foot electrode If the shoes are dissipative the measurement starts automatically After a short measuring time the r...
Страница 9: ...F OFF 20 M for single shoe 40 M for series OFF ON 35 M for single shoe 70 M for series ON OFF 70 M for single shoe 140 M for series ON ON 100 M for single shoe 200 M for series Switch 5 Lower limit OF...
Страница 10: ...F 3 value left shoe STX RSL SP value in kOhm UNR under range OVR over range ETX CR LF 4 value right shoe STX RSR SP value in kOhm UNR under range OVR over range ETX CR LF 5 test OK STX ERG SP OK ETX C...
Страница 11: ...e to low or to high 128 tbd 10 contact electrode was released to early If several errors occur the error numbers are added Example 1 Data STX RHG 2671 ETX CR LF STX ERG OK ETX CR LF Sense Resistance w...
Страница 12: ...serial port are located on the rear side of the unit Use a RJ12 western modular plug to connect the dry contact of the door opener The door opener relay is triggered and stays for 1s or 3 seconds when...
Страница 13: ...ated on the rear side of the unit Use a RJ45 western modular plug to connect PGT 120 COM to PC or terminal Pin 4 GND Pin 5 RxD Pin 6 TxD Connector View COM 8 7 6 5 4 3 2 1 The names RxD and TxD are re...
Страница 14: ...grease 2 Remove protecting foil of the velcro tapes and do not touch the sticky side 3 Apply the velcro tapes according to the picture onto the mounting plate 4 Remove the second protecting foil of t...
Страница 15: ...d with the instrument Serial port RS232 9600 baud 8 Data bit no Parity 1 Stop bit Measuring ranges Wrist strap 750k 35M Footwear each shoe 100k 100M Footwear in series 200k 200M Hands free Mode Tolera...
Страница 16: ...PGT 120 Personnel Grounding Tester LED upper limit exceeded LED below low limit LED measured value in the limit footwear test male snap 3mm below low limit LED upper limit exceeded LED wrist strap te...
Страница 17: ...09 07 socket for door opener sockets for foot electrode socket for external power supply Dip Switch socket for RS232 AC 12 V 7 6 5 2 8 9 10 1 OFF ON mode voltage beep series door opener high limit ser...
Страница 18: ...en vorbehalten Subject to change without notice WOLFGANG WARMBIER GmbH Co KG Untere Gie wiesen21 D 78247 Hilzingen Telefon 49 77 31 86 88 0 Telefax 49 77 31 86 88 30 e mail info warmbier com http warm...