MN-111 | Service book | Errors and Troubleshooting MN-111 ENG Rev08 | 2019/06/21 | © 2019 W&H Sterilization Srl
Other problems
WATER TANK PROBLEMS
Description
Possible causes
Checks and actions
Overflow of the clean water tank during filling (no
acoustic signal)
Upper clean water tank level sensor faulty
Check the sensor placement (must be horizontal).
Check the floating part moves freely.
Check the sensor connections and wiring.
Check electrical operation of the sensor. (*)
Overflow of the used water tank
Used water tank level sensor faulty
Check the sensor placement (must be horizontal).
Check the floating part moves freely.
Check the sensor connections and wiring.
Check electrical operation of the sensor. (*)
Water on the top lid or from the top cap, no warning
on screen)
The "Drain used water tank." message is displayed,
in spite the tank is empty.
Used water tank drains slow or not at all
Clogged tank outlet filter or clogged drain tube
Clean the used water tank outlet filter.
Check the tubes from the used water tank outlet to the drain
connection; clear or replace any clogged tubes.
Recommend the user to properly clean the load prior to sterilization.
Note
: If the sterilizer is not used for over a week, both tanks must be
emptied to avoid algae growth.
No water filling from the automatic system
Low-level clean water sensor faulty
Check for proper sensor placement.
Check the floating part moves freely.
Check the sensor connections and wiring.
Check electrical operation of the sensor. (*)
Water filling valve faulty
Check electric operation of the valve.
Check the valve connections and wiring.
The "Fill clean water tank." message is displayed, in
spite the tank is full.
Low-level clean water sensor faulty
Check for proper sensor placement.
Check the floating part moves freely.
Check the sensor connections and wiring.
Check electrical operation of the sensor. (*)
Note (*): see Test instructions service book.
WATER SPILLAGE PROBLEMS
Location of the water
Possible Cause
Checks and actions
Below the door
Dirt or debris under the door seal or chamber flange
Execute the monthly maintenance described in the Instructions for
use.
Improper door adjustment
Execute the Door adjustment cycle and procedure.
Door seal damaged
Replace the door seal.
Below the service door
Leak from the front drain connections
Check connections and tubes.
From the bacteriological filter
Leak of the check valve inside the chamber
Check the check valve inside the chamber behind the steam diffuser
Under the sterilizer
Leak from the water tank
If necessary replace the water tank.
Leak from the tank filter hose
Check connections and tubes.
If necessary replace tank filter hoses.
Leak from any of the water tubes (drain, water
pump feed)
Check connections and tubes.
Replace tubes.
PRINTER PROBLEMS
Description
Possible Cause
Checks and actions
Thermal printer
No report printed
Connection problem /
printer power problem
Check the printer serial cable.
Check the printer power supply.
Check (connect or replace) the serial (printer) flat cable coming off the
MMI board.
No paper
Check presence of paper.
Printer fault
Replace printer.
Wrong printer model selected
Set the right printer model. Please refer to the Instructions for use.