Maintenance
74
Lyla | Instructions for Use | RIN-210 ENG Rev02 | 15/09/2022 | © W&H Sterilization Srl
CLEAN THE WATER TANKS
Notice
: Do not touch the water level sensors. If misplaced or misaligned from their original position, the operation of the sterilizer could be
impaired.
Proceed as follows:
1
Clean the internal tank surfaces with a soft sponge moistened with water. For
better cleaning results, clean with W&H MC-1000 cleaning solution. Use a small
non-abrasive brush to clean the areas that are difficult to reach.
Notice
: never use any other disinfectant, detergent or abrasive product, as they
might result aggressive for the tank material.
2
Rinse the internal tank surfaces thoroughly, until all residuals of dirt and
detergent have been eliminated.
3
Dry the internal tank surfaces.
4
Detach the drain tube, attach it to the drain connection of the other tank and
repeat steps 1, 2, 3.