Sterilization cycles
END OF A STERILIZATION CYCLE
When a cycle is successfully finished, the "Cycle completed"
message appears on the screen. To end the cycle:
1
Tap
to view the cycle
summary or tap
to view the
cycle parameters. See "View
the cycle parameters" on the
previous page.
2
Tap OPEN to open the door: the
door unlocks and the home
page appears.
Note: if an error message appears see
"Troubleshooting" on page 106
CAUTION! Hot surfaces.
Burnings. Do not touch the
chamber, the internal side
of the door and the internal
fittings. Use the tray holder
or cassette holder or
gloves for high
temperatures or adequate
protection to remove the
load!
3
Open the chamber door.
4
Remove the load and stock it.
STOP A STERILIZATION CYCLE
WARNING! You can stop the cycle at any time. Instruments
must not be considered sterile if this occurs before the DRY
phase.
A cycle can be manually aborted at any time. To stop a cycle:
1
Tap STOP: a confirmation request
appears.