50
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
Sterilization cycles
Standard version
269.6 °F (132 °C) cycles
249.8 °F (121 °C) cycles
Duration
(minutes)
4
30
Min.
temperature
269.6 °F (132 °C)
249.8 °F (121 °C)
Max.
temperature
275 °F (135 °C)
255.2 °F (124 °C)
Min. pressure
26.68 psi at 269.6 °F
(1.84 bar) at (132 °C)
14.9 psi at 249.8 °F
(1.028 bar) at (121 °C)
Max. pressure
30.45 psi at 275 °F
(2.1 bar) at (135 °C)
17.83 psi at 255.2 °F
(1.23 bar) at (124 °C)
DELAY THE STERILIZATION CYCLE START
You may schedule the start of the sterilization cycles at a certain
date and time (e.g., if you want to load the sterilizer in the evening
and run standard sterilization cycle early the next morning before
office hours). You can set the cycle start date and time and enable or
disable it for each cycle.
1
On the homepage, tap the cycle and
.
2
Tap
Start cycle at
.
3
If you want to change the start time, tap the time or the date: a
settings page opens.
4
Tap the number you want to change and tap
or
to
increase it or decrease it.
5
Tap
to confirm. This date and time become the scheduled
default start time for all next sterilization cycles.
6
Tap
to lock the door; a new page appears.
Note
: if nothing else is pressed, the cycle will start at the programmed
time. The page allows also to start the cycle immediately ("Start now") or to
delete the operation and the programmed cycle ("Stop").
SET THE DRYING TIME
Drying time is set by default for each sterilization program. The
default drying time for each sterilization program can only be
increased from the default value.
1
On the homepage, tap the cycle and
.
2
Tap the drying time: a settings page opens.