Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
79
Maintenance
4000 cycle or five-year maintenance
GENERAL CHECK AND SERVICE REQUIRED
Notice
: Regular service is imperative to ensure continuous and
effective operation of the sterilizer.
A general check and service should be carried out every 4000 cycles
or five years by an authorized service technician. The service
required includes the following:
n
the replacement of consumables and other important internal
components
n
a check of the entire sterilizer with special care for the safety
systems
n
the cleaning of areas and components that cannot be accessed
by the user.
ACTIONS REQUIRED FOR EACH ELEMENT
For each element, the actions to carried out are the following:
Element
Replace
Clean
Check
Electro valves
x
-
-
Vacuum pump internal parts
x
-
-
Sterilization chamber and external surfaces
-
x
-
Chamber filter
-
x
-
Element
Replace
Clean
Check
Internal parts, with particular care for the
condenser fins and the main board
-
x
-
Pneumatic connections
-
-
x
Electrical connections
-
-
x
Temperature and pressure calibration
-
-
x
Door locking system
-
-
x
Pressure safety valve
-
-
x
Safety systems
-
-
x