(3) External PAR sensor not connected or incorrect PAR chan-
nel selected in window “External PAR Sensor” (Fig. 52,
page 89).
PAR sensor compatible with MNI-PAM-II: To calibrate the inter-
nal PAR sensor using the PAR sensor of the 2035-B clip con-
nected to the MINI-PAM-II, go to “Calibrate” and establish cali-
bration factor automatically.
Other PAR sensor: To calibrate the internal PAR sensor using an
external PAR sensor which cannot be read by the MINI-PAM-II,
adjust calibration factor manually by selecting “Calibration Fac-
tor”, pressing SET and changing the factor by arrow keys until
internal PAR matches that of the external sensor.
Leaf Clip/Ext. PAR
7.3.4.2
Ext PAR Sensor of the menu “Leaf Clip/Ext. PAR” opens a sub-
menu that selects the active external PAR channel (Fig. 52, page
89). Four possible PAR sensors are considered: PAR sensors 1
and 2 of 2035-B leaf clip, the miniature spectrometer, and a PAR
Leaf Clip/Ext. PAR Sett.
SET
EXIT
Side Panel
,
. Up/down
keys move cursor.
SET
. Open submenu
Calibrate.
EXIT
. Return to previous
menu.
Ext. PAR Sensor
Leaf Clip Ch. 1
Leaf Clip Ch. 2
Temperature
Fig. 51:
Leaf Clip/Ext. PAR Sett.