Chapter 7
Touchscreen Operation
66
In the Recovery window, the MEM button opens the window
“Recovery Memory” in which all recovery kinetics can be viewed
independent if they are connected to an induction or light curve
or if they represent separate experiments.
Table 7: Time Sequence of Saturation Pulse Analyses in a Recovery Curve
SAT number
Time in darkness, min
1
0.00
2
0:30
3
1:30
4
4:00
5
9:00
6
19:00
7
39:00
7.2.10 Actinic Light List
The window “Actinic Light List” contains PAR values (in
µmol m
-2
s
-1
) which increase roughly exponentially with intensity
settings 1 to 12. The entire PAR list can be multiplied by the in-
ternal actinic factor (“Int. Act. Factor, Fig. 41, page 77).
In the factory, the internal actinic light is adjusted so that the PAR
values of the Actinic Light List apply to the sample level of the
2035-B leaf clip with the MINI-PAM-II fiber optics fully inserted
(i.e. distance between fiber optics tip and sample level of 7 mm;
angle between end piece of fiber optics and sample level of:
60°). Because of similar geometrical arrangement, these factory
settings are also valid for the 2060-B Arabidopsis Leaf Clip when
distance rings 2 mm plus 4 mm are used, and the 2010-A Dis-
tance Clip 60°.