User Interface 

29 

Waltron User Manual 

9065 Luminescent Dissolved Oxygen Analyzer 

 

Table 4.3: Configurable Alarm Settings. 

 
 

4.5.11 

ALARM OUTPUTS 

An overview is given below of the alarm outputs and the corresponding activation conditions. 
See Section 3.6.5 for an overview of the physical I/O connections.  
 

General Alarm (A1): 

 

All faults. 

 

All error messages. 

 

Temperature alarms for sensor 1 and/or sensor 2. 

 

O

2

 alarms for sensor 1 and/or sensor 2. 

Digital output A1 (Normally Close, NC) is zero if there has been an alarm during two 
measurements. The alarm can be reset by selecting ‘ACK/RESET ALARM’ in the Shortcut 
menu. The digital output is then high (and remains high during the alarm signal). 

 

O

2

 Alarm (A2): 

If the O

2

 value of sensor 1 and/or sensor 2 is greater than the set value, then output A2 is 

zero. N.B. output 1 is also zero in the case of an O

2

 alarm. 

These outputs can be reset by selecting ‘ACK/RESET ALARM’ in the Shortcut menu and 
will then remain high (if the alarm signal continues). 

 

O

2

 Extra (Alarm) for Sensor 1(A3): 

If the O

2

 value of sensor 1 is greater than this value, then output A3 is zero. 

For example, in the case of filtration, if the value is greater than the O

2

 alarm high-high value 

(O

2

 extra), then output A3 cuts out. As a result, the process of the sensor 1 circuit can be 

stopped.  Or, for example, if the value is less than the O

2

 low alarm value (O

2

 extra), then 

output A3 is activated. As a result, the aeration of the sensor 1 circuit can be started. 

 

O

2

 Extra (Alarm) for Sensor 2 (A4): 

If the O

2

 value of sensor 2 is greater than this value, then output A4 is zero. 

The same conditions apply here as for the O

2

 extra alarm for sensor 1. 

 
 
 
 
 
 
 
 

Item 

Range 

Default 

Unit 

O

2

 alarm 

0 …. 45,000 

100 

ppb; ppm; mg/l; ug/l; % as; %  

O

2

 extra 

0 …. 45,000 

200 

ppb; ppm; mg/l; ug/l; % as; %  

Low temperature 

10 … 150 

20 

° C; ° F 

High temperature 

10 … 150 

20 

° C; ° F 

Содержание 9065

Страница 1: ...Revision 2 05 EXPERTS IN WATER CHEMISTRY SINCE 1903 9065 Luminescent Dissolved Oxygen Analyzer User Manual...

Страница 2: ...4 5546 www waltron net Please be ready to provide the following information Date analyzer was purchased Analyzer model and serial number Recent maintenance history Calibration slope values and detaile...

Страница 3: ...containers packages and chemicals Only qualified personnel should be involved in the installation operation and servicing of the analyzer Follow safety precautions when operating analyzer in condition...

Страница 4: ...chnical Service Department Customer name address and department Name and telephone number of the individual responsible for returning items for repair Brief problem description Ship to Waltron service...

Страница 5: ...moving and shipping needs Check all items received against those on the packing list Chemicals are usually shipped in a separate package and will be itemized accordingly Verify that the number of pack...

Страница 6: ...2 1 Analyzer Overview 12 2 2 Sensor 13 2 3 Coatingholder 13 2 4 O2 Measurement Principle 13 3 Installation 15 3 1 Content of the Delivery 15 3 2 Pre Installation 15 3 3 Mounting the 9065 16 3 4 Assem...

Страница 7: ...on of the Temperature Sensor 33 6 2 Calibrating the O2 Sensor 33 6 2 1 Calibration Frequency 33 6 2 2 One point Calibration 33 6 2 3 Calibration with Two Gases 36 6 2 4 Check Calibration 38 6 2 5 Cali...

Страница 8: ...6 Figure 3 2 Dimensions of the 9065 sensor mounting bracket 16 Figure 3 3 Fitting the sensor into the flowcell 17 Figure 3 4 Sensor orientation relative to flow direction 17 Figure 3 5 Sample Connecti...

Страница 9: ...ble 3 3 Connector 1 Power supply 24V DC 19 Table 3 4 Connector 2 RS 232 communication 20 Table 3 5 Connector 3 and 4 Sensor 1 and 2 Sensor communication 20 Table 3 6 Connector 5 I O Cable 20 Table 4 1...

Страница 10: ...ed value ppm 0 001 ppm 2 of the measured value mg l 0 001 mg l 2 of the measured value Temperature C 0 1 C F 0 2 F Max Pressure barg 10 00 barg psi 145 0 psi kg cm2 10 20 kg cm2 kPa 1 000 kPa Measurem...

Страница 11: ...er can be used in a variety of online analysis applications throughout many different industries The 9065 must be connected to a sample line The connections to the flowcell are 6mm Swagelok fittings T...

Страница 12: ...ECAUTIONARY MEASURES AND SAFETY INSTRUCTIONS WARNING To avoid a short circuit never insert metal objects into the connector The use of chlorite based or fluorine based cleaning products is not permitt...

Страница 13: ...lved Oxygen Analyzer 2 INTRODUCTION 2 1 ANALYZER OVERVIEW The 9065 has two main sections 1 Transmitter 2 Sensor Housing Figure 2 1 Analyzer overview Transmitter Display User Interface Transmitter Conn...

Страница 14: ...of the measured partial oxygen pressure and the temperature The oxygen sensor optically measures the liquid s O2 content based on the luminescence measurement principle where an oxygen sensitive layer...

Страница 15: ...oxygen 1 and in the presence of oxygen 2 The oxygen content is calculated using the time difference between the exposure to the blue light and the molecules lighting up phase shift and the product te...

Страница 16: ...et Always state the serial number the order number or the invoice number as given by Waltron of the 9065 in all correspondence NOTE Before using the instrument you must make sure the instrument is com...

Страница 17: ...alyzer 3 3 MOUNTING THE 9065 The 9065 can be mounted on a wall or a flat surface using the dimensions shown in the figure below Figure 3 1 Dimensions of the 9065 mounting plate for wall mounting Figur...

Страница 18: ...plate The figure below shows how the sensor properly fits into the flowcell It is important that the coating is always the first to be in contact with the flowing product The sensor can be located a...

Страница 19: ...inlet and outlet sample connections to the flowcell are Swagelok fittings It is highly recommended to us stainless steel tubing for the sample lines to prevent outside oxygen from entering the sample...

Страница 20: ...ector No Cable Connection 1 Power supply 2 RS communication 3 Sensor communication 1 4 Sensor communication 2 5 I O communication Profibus Depending on model 3 6 2 POWER CABLE Table 3 2 Connector 1 Po...

Страница 21: ...Txd Grey 6 NTC An inp Pink 7 NTC common Blue 8 Shielding 3 6 5 I O CABLE Table 3 6 Connector 5 I O Cable Pin No I O Type Comment Wire color 1 Ground Ground White 2 24V DC Supply Brown 3 24V control A...

Страница 22: ...n example of the I O connections 3 7 PREPARATION AND START UP Power can be supplied to the analyzer afternall the cables have been connected The 9065 will initialize automatically once the power is tu...

Страница 23: ...tton is pressed for 3 seconds whilst the measurement screen is displayed then the main menu screen will be displayed Up Arrow For a quick overview the up arrow button can be used to scroll to the SHOR...

Страница 24: ...ors are connected and one or both do not give a measurement value then a dotted line will be displayed on the screen and the status message will change The current time and the operating mode remote l...

Страница 25: ...ibration mode is activated During calibration mode it is not possible to take a measurement 4 4 USER LEVEL The default settings can be changed to the preferred user settings before the 9065 is used fo...

Страница 26: ...e Configuration menu 1 In the main measurement menu press the Enter button for three seconds The main menu will be displayed 2 Use the cursor keys to select Configuration and press the Enter button to...

Страница 27: ...Configuration menu 4 5 5 LANGUAGE The arrow buttons can be used to select Dutch German or English as the language Press the Enter button to confirm the selected language The Configuration menu will th...

Страница 28: ...be compensated with the measured pressure For example Patm during calibration 1013 mBar O2 measured 1 033 Bar 20 cm H2O 1 0 O2 Patm 0 981 4 5 8 2 Liquid Application When the sensor will be used in liq...

Страница 29: ...e setting concerned 4 Press the Enter button to return to the Analog outputs menu 5 Use the cursor keys to select the Return sign and press the Enter button to return to the Configuration menu NOTE Th...

Страница 30: ...1 is also zero in the case of an O2 alarm These outputs can be reset by selecting ACK RESET ALARM in the Shortcut menu and will then remain high if the alarm signal continues O2 Extra Alarm for Senso...

Страница 31: ...Summary x The output becomes zero immediately the relay cuts out due to the alarm function The state of the relay remains unchanged Sensor Alarm A1 A2 A3 A4 1 Temp min max X 1 O2 O2 alarm X X 1 O2 O2...

Страница 32: ...ull the oldest measurements are deleted and the most recent measurements are saved see Section 5 4 5 3 LOCAL AND REMOTE OPERATION Before the 9065 is used for the first time all of the desired paramete...

Страница 33: ...e sensor are displayed A distinction is made between measurement hours and standby hours Extremes The extreme values for the oxygen content and the temperature are saved in this menu Calibration The l...

Страница 34: ...se counter will be reset and the warning symbol will no longer be displayed The following two principle methods are described to calibrate the oxygen sensor 1 One point calibration This method is reco...

Страница 35: ...ate O2 calibration the sensor must be at the same temperature as the gas To achieve this allow the sensor and calibration gas to acclimatize for 4 hours in the same room Procedure 1 Put analyzer into...

Страница 36: ...alibrated Note that a single stream analyzer will only display Sensor 1 as an option 11 Select O2 Auto and press enter 12 Select O2 Low and press enter 13 Press Enter again and calibration is now in p...

Страница 37: ...ect the back arrow to return to the previous menu Repeat until arriving at the main measurement screen The calibration process is complete 22 Disconnect the calibration gas from the calibration beaker...

Страница 38: ...the a s unit NOTE For an accurate O2 calibration the sensor must be at the same temperature as the gas To achieve this allow the sensor and calibration gas to acclimatize for 4 hours in the same room...

Страница 39: ...rocess is complete 32 Disconnect the calibration gas from the calibration beaker and remove the calibration beaker from the sensor 33 Replace the flowcell on the sensor again taking caution to prevent...

Страница 40: ...Enter button to confirm the selection 5 In this screen the analog outputs can be calibrated to the correct mA value A milliammeter or a multimeter must be connected to the appropriate output to measu...

Страница 41: ...with a detergent to clean the sensor or the outside of the control box Warning Never use tools to clean the sensor coating Touching the coating may cause serious damage to the instrument Regular CIP o...

Страница 42: ...ge the high level O2 high high O2 value set O2 high high alarm Check the process Change the high high level Sensor x Temp high alarm Temp value set Temp high value Check the process Change the Temp hi...

Страница 43: ...G HOLDER If the coating is faulty or damaged then a new coating holder must be placed on the sensor The instructions for replacing the coating are given below 1 Remove the sensor from the mounting pla...

Страница 44: ...side of the coating holder s screw thread This is to stop the coating holder from shaking loose Make sure no glue gets on the pink side 6 Screw the coating holder back on to the sensor return the sens...

Страница 45: ...M 9 2 STORAGE The instrument must be stored in a dry location and be protected against mechanical and or chemical damage If the 9065 is not going to be used for a long time the instrument must be thor...

Страница 46: ...for 24V DC power supply 3p male K5000 015 Power cable wide range 110 240 VAC 5 m K5000 005 Power cable wide range 110 240 VAC 15 m K5000 016 Power cable wide range 110 240 VAV 40 m K5000 017 Power cab...

Страница 47: ...lyzer LDO sensor K5000 001 Calibration set K5000 028 Flowcell K5000 002 CD rom with PC software for reading data to PC Incl RS 232 cable On request Swagelok connector P1000 106 25 mm Orbisphere adapte...

Страница 48: ...9065 can be set to various oxygen units where the instrument then makes the following conversions 100 20 95 saturation air 2 O With Tm Temperature in C patm Atmospheric air pressure during the oxygen...

Страница 49: ...DIX B SOFTWARE STRUCTURE Key Start screen Main menu Shortcut Key Automatic after 3 sec Sensor 2 Key Sensor 1 Sensor 1 2 Measurement screen key Key O2 trend graph Sensor 1 Measurement values Sensor 1 D...

Страница 50: ...alibration sensor 1 B Calibration sensor 2 B A Operator level B Supervisor level C Service engineer level Main menu Configuration A Info A Units B Sensor settings B Analogue outputs B Alarm settings B...

Страница 51: ...B Calibrations A History 1 Sensor1 A History 2 Sensor1 A History 3 Sensor1 A History 1 Sensor2 A History 2 Sensor2 A History3 Sensor2 A Display temperature graph s2 A Display O2 graph s2 A Display tem...

Отзывы: