background image

 

1.0  Introduction 

 

The Walchem disinfection sensors consist of an amperometric 
sensor assembly and a flow cell. Assembly of these parts is 
required, so please read these instructions carefully. The sensor is 
capable of measuring the disinfectant in clean water or in water 
contaminated with debris thanks to our unique flow cell design. 
 
WDSW100, WCNP100, WDS600, W900 and WebMaster series 
controllers with amperometric chlorine sensors can be used for 
reporting chlorine residual measurements in accordance with EPA 
Method 334.0.

 

 
The standard (part numbers 104010

,

 

191300, 191441, 191530 & 

191587) free chlorine/bromine sensor membrane is not compatible 
with water containing surfactants! 

 

Sensor 

The sensor assembly includes the sensor body, a replaceable 
membrane cap, a 100-ml bottle of electrolyte fill solution, and 
special abrasive paper. Make sure that all parts are included. 
 
The oxidizer molecules diffuse through the membrane and in the 
acidic environment of the electrolyte fill solution, a redox reaction 
occurs at the electrodes in the sensor. The current generated by this 
reaction is converted to a robust voltage signal that is linear with the 
concentration of the oxidizer. 
 

Flow Cell 

The flow cell consists of a translucent flow cell body, mounting nut 
and o-ring, washer set and o-ring. Make sure that all parts are 
included.  
 
The flow cell is required to prevent bubble formation on the 
membrane and to provide proper flow velocity across the face of the 
membrane. The sensor will not read accurately if it is not installed 
in the flow cell, with a flow rate between 30 and 100 liters per hour, 
at an operating pressure of 1 atmosphere or less. The ¼” adapter 
must NOT be removed from the flow cell. 

Содержание WDSW100

Страница 1: ...Five Boynton Road Hopping Brook Park Holliston MA 01746 USA TEL 508 429 1110 WEB www walchem com Disinfection Sensors Instruction Manual...

Страница 2: ...chanical parts and electrodes from date of delivery from the factory or authorized distributor under normal use and service and otherwise when such equipment is used in accordance with instructions fu...

Страница 3: ...ions 8 3 0 Operation 13 Conditioning 13 Calibration 13 4 0 Troubleshooting 15 The disinfectant reading is much lower than the manual analysis 15 The disinfectant reading is much higher than the manual...

Страница 4: ...embrane cap a 100 ml bottle of electrolyte fill solution and special abrasive paper Make sure that all parts are included The oxidizer molecules diffuse through the membrane and in the acidic environm...

Страница 5: ...he rubber band from the groove in the membrane cap just until the vent hole underneath is exposed then fill the membrane cap to the top with the electrolyte fill solution Never shake the electrolyte b...

Страница 6: ...peat several times at different angles Never touch or clean the brown electrode shaft 5 Holding the electrode shaft upright place it onto the filled membrane cap with the mounted G holder You may have...

Страница 7: ...e that the bands are smooth and flush 5 Rinse your hands the sensor and all surfaces contaminated with electrolyte solution with running water Check the sensor for leaks especially at the membrane and...

Страница 8: ...uld be installed in an area where there is good solution movement and where it will respond rapidly to chemical additions The placement of the sensor relative to the placement of chemical replenishmen...

Страница 9: ...iew Disinfection Sensor Washer Set p n 103419 O Ring p n 103422 Nut p n 102586 Cable p n 191303 for WDIS400 WebMaster Cable p n 191655 for W100 W600 Membrane cap Clip Ring groove Clip ring O Ring p n...

Страница 10: ...Typical Installation ISOLATION VALVE NORMALLY OPEN RECIRCULATION PUMP ROTAMETER 30 100 LPH PROCESS WATER SAMPLE RETURN 1 ATMOSPHERE MAXIMUM FLOW SWITCH SENSOR FLOW CELL FLOW CONTROL VALVE SAMPLE VALV...

Страница 11: ...le The wiring to the controller is as follows Shield Drain Earth Ground GRN IN WHT IN RED 5 V BLK 5 V If the required cable length exceeds the 6 meters 20 feet that is supplied wire the housing to a 1...

Страница 12: ...the instrument The maximum cable length is 30 5 meters 100 feet WDIS410 Power Supply 115 VAC or 230 VAC Contact Closure Polarity not critical Interlock Function 5V T T GROUND STUD L1 L2 N G R N 1 2 0...

Страница 13: ...2 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 TB1 TB4 TB6 TB5 TB7 N N N N N N R1 R2 R3 R4 R5 R6 POWER SWITCH...

Страница 14: ...4 5 6 7 8 9 10 11 12 TB2 TB1 FUSE POWER SW ITCH POWER SWITCH 1 2 3 4 5 6 7 8 9 10 11 12 R1 R2 R3 N N N N L TB3 TB4 TB5 TB6 TB7 pH TB1 1 2 3 4 5 6 7 8 9 10 11 12 5V BLK IN WH T IN GRN SHIELD 5V RED SAF...

Страница 15: ...12 W900...

Страница 16: ...f the application 4 The stability of the sensor and controller as a system The frequency of calibration is really determined by experience At a new installation calibration might initially be checked...

Страница 17: ...14...

Страница 18: ...ions 5 When the reading is stable continue to the final steps of the calibration NOTE Disinfectant concentration can change rapidly in the sample Minimize the time between performing the DPD test or m...

Страница 19: ...ring Check wiring Faulty sensor Replace sensor Faulty controller sensor input Go to the Sensor Input menu and perform a self test If this passes then the problem is with the sensor or its wiring If it...

Страница 20: ...or the appropriate amount of time before attempting a calibration Insufficient sample flow Increase flow rate to between 30 and 100 liters per hour Air bubbles on membrane Dislodge bubbles Adjust flow...

Страница 21: ...ctants on the standard free chlorine bromine sensor membrane as these will reduce the life of the membrane If the sensor still cannot be calibrated after cleaning replace the membrane cap as described...

Страница 22: ...ubmerged in water For long term storage up to 3 years follow this procedure CAUTION Wear gloves and safety glasses during assembly of the sensor since the electrolyte is a STRONG ACID It is recommende...

Страница 23: ...erating Pressure Sensor 0 to 3 atmosphere 0 to 44 psi Total Chlorine CP4 0 to 1 atmosphere 0 to 14 7 psi All other sensors Operating Pressure Optional flow switch 0 to 10 3 atmosphere 0 150 PSI see gr...

Страница 24: ...30 to 100 liters hour 0 13 to 0 44 gal min pH Range 6 8 8 0 pH must be stable within 0 10 Conductivity Range Up to 4 NaCl Response time 30 sec Conditioning time 60 min Free Chlorine Bromine Extended p...

Страница 25: ...l 0 01 mg l 0 200 mg l 0 167 5 mg l 0 100 mg l 0 1 mg l Sensitivity Free Chlorine 5 Ozone 2500 Sample Flow rate 30 to 100 liters hour 0 13 to 0 44 gal min pH Range 1 0 11 0 Conductivity Range No Limi...

Страница 26: ...me 60 min Hydrogen Peroxide Range W100 W600 W900 Range WDIS Range WM1 WIND Resolution 0 200 mg l 0 167 5 mg l 0 100 mg l 0 1 mg l 0 2 000 mg l 0 1 675 mg l 0 1000 mg l 1 mg l 0 20 000 mg l 0 16 750 mg...

Страница 27: ...SOR ClO2 2 PPM 4 20 mA 191648 103462 103475 SENSOR ClO2 5 PPM 4 20 mA 191650 103462 103475 SENSOR H2O2 200 PPM WP7 191537 103462 103761 SENSOR H2O2 2000 PPM WP7 191539 103462 103761 SENSOR H2O2 20000...

Страница 28: ...25...

Отзывы: