Wahl DSTCAL Software User Manual
12
8. Software will d
isplay “Probe Calibration Success” window.
9. Click OK
10. Software will return to main screen and new value will be loaded into
the R0 register and window.
4.4. Error Windows
The following error box may occur when in the Meter Only Calibration routine and if the
input resistance is not within approximately 4 ohms of the requested nominal setting.
One or both of the following error messages may occur when in the Meter Only
Calibration and the calculated calibration values don’t meet the expected results. The
most likely cause of this is that the resistances applied were not equal to the specified
resistance or the resistance was changed during the calibration measurements. This
may be corrected by entering the actual resistance of the standard used.