background image

10/01/13 WD1100 Rev A 

 

 

 

Digi-Stem

®

 Thermometer 

 

Panel Mount Installation Instructions 

 

The Panel Mount system for Wahl Digi-Stem

®

 is designed for use with panels 

from 1/8” to 3/8” thickness.  The system uses a square O-ring to seal around the 
meters H-Frame and the panel.  The unit is secured by a U-bracket with 2 #8 
wing-nuts and lock washers.

 

 
For Panel cutout dimensions and installation reference Wahl Drawing 13615. 
 
1) Remove Hardware and bracket from mounting studs on rear of meter 
enclosure. 
 
2) Ensure the square O-ring is installed flat against the H-frame and is not 
twisted. 
 
3) Slide the meter from the front surface of the panel through the panel opening.   
 
4) Install the U bracket over the mounting studs with the short arms facing the 
rear side of the panel. 
 
5) Install a lock washer and wing-nut onto the studs. 
 
6) Gently tighten the wing-nuts until they just start tightening.  The bracket should 
not be deflected more than 0.020” (approximately 

¾

 to 1 turn after touching).    

 

CAUTION:

 Excessive force may result in bending and fracturing the U bracket or 

damaging the mounting studs. 
 
 
 
 
 
 

Wahl Instruments, Inc. 

234 Old Weaverville Road 

Asheville, NC 28804 

Toll Free 800-421-2853 

Phone 828-658-3131 

Fax 828-658-0728 

www.palmerwahl.com

 

 