 CONTROL CENTER - TEMPERATURE MONITOR SYSTEM  

 
The  temperature  control  center  (monitor)  system  on  the  front  of  the  unit  provides  the 
following functions- 

-

 

 to provide an indication of the sample temperature inside of the compartment  

-

 

to provide a local alarm at the unit in the event an out of range temperature exists 

-

 

to  provide  the  ability  to  connect  remote  systems  to  provide  remote  alarms  or 
notifications in the event of an out of range temperature.  

-

 

to provide the ability to adjust setpoints, as well as program and test other parameters 
associated with the alarm system (i.e. delay times, system status). 

-

 

a battery backup, which will maintain the alarm system for several hours after a power 
failure to allow for continued alarm and notification functions.  

The temperature monitor system does NOT control the temperature of the unit

. That function 

is performed by the Microprocessor Digital Temperature Control, which is preset at the factory 
and should not need to be adjusted by the user.  

 

 

 

C

ONTROL 

C

ENTER 

P

ANEL

 

The LCD Display shows the temperature in ºC  

 

In  the  event  of  a  loss  of  power,  the  display  shows  a  BAT  message  to  indicate  that  the 
monitor

 

is running on the battery.  If the battery is low, the BAT message will blink. 

 

 

ALARM 

MUTE/PROGRAM  

UP button 

DOWN button 

STATUS tri-color LED 

ALARM/TEST button 

 

Alarm beeper 

Содержание SCCP-19

Страница 1: ...agent driver at time of delivery Make sure you get a signed copy Send a written request for an inspection to the carrier B Concealed Damage This type of damage may not be discovered until the unit is...

Страница 2: ...MANUFACTURER S TECHNICAL SERVICE DEPARTMENT please have the unit s model and serial number ready as well as the problem description The model and serial number is located on the serial tag which can b...

Страница 3: ...y receipt acknowledging the damage This has to happen at the time of delivery or it won t happen at all Keep a copy for your records and send another to the carrier s damage claims department along wi...

Страница 4: ...t needed PLEASE NOTE Your refrigerator or freezer is designed as INDOOR USE and should be operated at air conditioned room temperature between 65 F to 85 F Remove the lower grill assembly from the fro...

Страница 5: ...the plastic door stoppers see Image 4 5 and 6 CASTERS INSTALLATION It is important to ensure the casters are installed properly and securely in order to obtain the maximum strength and stability of th...

Страница 6: ...on cord Be sure your unit is properly grounded Use the 3 prong plug provided into a 3 prong grounded outlet Only this method complies with national electrical codes local codes and ordinances Unless t...

Страница 7: ...nd product temperatures inside of the units and require no further adjustment Please contact the manufacturer s Technical Support Department before making any adjustment to make sure the adjustment is...

Страница 8: ...king the evaporator fan s if installed There should be at least 4 inches of clearance below the fan s to allow proper air flow Do not overload the unit Maximum suggested load is about 75 of chamber ca...

Страница 9: ...ated with the alarm system i e delay times system status a battery backup which will maintain the alarm system for several hours after a power failure to allow for continued alarm and notification fun...

Страница 10: ...pressed The MUTE PROGRAM button has three functions In Normal Run mode if the unit is beeping pressing the button will mute the beeper for a preset time Each time the button is pressed the mute time r...

Страница 11: ...the BEEPER and ALARM LED will both activate Press the MUTE PROGRAM button to stop the beeper for a specified time changing the sound to short chirps When the temperature goes back within range and the...

Страница 12: ...n a short period 5 Once the Monitor is in Programming mode each time the Mute button is pressed the display steps to the next parameter As the display steps to the next parameter the value of the prev...

Страница 13: ...mote Alarm Relay is a single pole double throw SPDT relay that changes state switches with the loss of line power or with the presence of an alarm condition NOTE The relay is rated for pilot duty oper...

Страница 14: ...of the unit A touch keypad allows the user to easily select the display units set point differential set point and heating or cooling mode Please Note The digital temperature controller has been fact...

Страница 15: ...et point and S1 will be blinking Press the up and down keys to adjust to the desired setting S1 blinking Setpoint Temperature Press SET key again to access the differential mode The LCD will display t...

Страница 16: ...y EP Appears when the probe is open shorted or sensing a temperature that is out of range Check to see if the sensed temperature is out of range If not check for probe damage by comparing it to a know...

Страница 17: ...placements are pushed against rear wall or interrupted the proper refrigerator air circulation For the proper air circulation place the products evenly on each shelf Do not push against the refrigerat...

Страница 18: ...prevent leakage Door will not close The unit is not level Refer to the Leveling section at the beginning of this manual Check for dirt and debris or items passing through the door seal MOISTURE DURIN...

Страница 19: ...sehold cleaners ammonia or alcohol to clean handles Use a soft cloth to clean smooth handles Do not use a dry cloth to clean smooth handles For Glass Door Units clean the glass with a mild detergent a...

Страница 20: ...th pressurized air CO2 or nitrogen REMEMBER THE CLEANING OF THE CONDENSER IS NOT COVERED BY THE WARRANTY AND IS YOUR RESPONSIBILITY ANY DAMAGE CAUSED BY FAILURE TO KEEP THE CONDENSER CLEAN IS ALSO NOT...

Страница 21: ...Y In addition to the two 2 year warranty stated above Horizon Scientific Inc warrants its hermetically and semi hermetically sealed compressors to be free from defects in both material and workmanship...

Страница 22: ...od or other external causes It does not apply to defects resulting from failure to properly install operate or maintain the product in accordance with the printed instructions provided or damage cause...

Отзывы: