Transmitting data (USB interface)
MD 8000 L
80
ba77218e01
11/2018
13 Transmitting data (USB interface)
13.1 Outputting current measurement data
13.2 Transmitting data (to a PC)
The meter has a USB-B interface (
USB Device
) e.g. to connect a PC.
Via the USB-B interface (
USB Device
) you can transmit data to a PC or printer
and update the meter software.
13.3 Connecting the PC / USB-B interface
(
USB Device
)
Connect the MD 8000 L to the PC via the USB-B interface.
Installation of the
USB driver on the
PC
System requirements of the PC for installation of the USB driver:
PC with at least one free USB connection and CD-ROM drive
Windows 2000, Windows XP, Windows Vista or Windows 7.
1.
Output the current measurement data to the USB-B interface with
<F2>
[USB output]
.
1.
Insert the supplied installation CD in the CD drive of your PC.
2.
Install the driver from the CD.
Follow the Windows installation instructions as necessary.
3.
Connect the MD 8000 L to the PC via the USB-B interface.
The meter is listed as a virtual COM interface among the connections in
the Windows instrument manager.
4.
Set the same transmission data at the connected instrument (PC):
Baud rate: to be selected in the range 1200 ... 19200
Handshake: RTS/CTS
Set at the PC only:
– Parity: none
– Data bits: 8
– Stop bits: 2
Содержание MD 8000 L