12.2 Automatic data storage at intervals
The storage interval (
Interval
) determines the time interval between automatic
data storage processes. Each data storage process transmits the current data-
set to the USB interface.
Configuring the
automatic memory
function
Settings
You can configure the automatic data storage function with the following set-
tings:
1.
Press the
<STR
_
>
key.
The menu for automatic data storage appears.
1 Specified entire storing duration
2 Max. available storage duration
3 Graphical display of the storage usage
ID number
1
Interval 30
s
Duration 180
min
continue
0d03h00min
0
1d17h33min
Automatic data storage
1
2
3
01.11.2018
08:00
Back
Menu item
Possible setting
Explanation
ID number
1 ... 10000
ID number for the dataset series.
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min, 30 min,
60 min
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the stor-
age duration.
Duration
1 min ... x min
Storage duration.
Specifies after which time the auto-
matic data storage should be termi-
nated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.