MD 8000 L
Settings
ba77218e01
11/2018
67
switch to the measured value display with
<M>
.
Default settings are printed in
bold
.
Setting menu
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of
the last calibration.
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the last calibration
records (max. 10)
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the calibration data stor-
age to the interface
Calibration
/
Calibration interval
1 ...
150
...
999 d
Calibration interval
for the IDP
conductivity sensor (in days).
The meter reminds you to cali-
brate regularly by the flashing
sensor symbol in the measuring
screen.
Type
Used measuring cell
Cal
Measuring cells whose cell con-
stant is determined by calibra-
tion in the KCL control standard
solution.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant
is displayed in the status line.
man
Freely (manually) adjustable cell
constant in the range 0.450 ...
0.500 cm
-1.
Cell const. man
0.450 ...
0.475
...
0.500 cm
-1
Display and setting options for
the manually adjustable cell con-
stant.
This menu item is only available
when
Type
man
is set.
Temp. comp. (TC)
/
Method
nLF
Lin
off
Procedure for temperature com-
pensation (see section 9.2 T
51).
This setting is only available for
the measured parameters, con-
ductivity (
ϰ
) and resistivity (
ρ
).