
ORP
MD 8000 L
40
ba77218e01
11/2018
to measure and enter the temperature of the sample.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
7.2
ORP calibration
Temperature
sensor
Resolution of the
temp. display
Temp. measurement
yes
0.1 °C
Automatic with temperature
sensor
-
1 °C
Manual
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.
Содержание MD 8000 L
Страница 93: ...MD 8000 L Glossary ba77218e01 11 2018 93...
Страница 97: ......