25
vwr.com I Instruction manual VWR
®
CO
2
incubator
Segments edition
10.1.2.
Press the button
, and the first program segment will appear
.
In this window you can set:
temperature
– set temperature which the unit is to achieve in this segment, (can’t be higher than over temperature
protection -2°C),
time
– the time of maintaining the set temperature ([d hh:mm]) (days, hours and minutes) it is possible to select
continuous work in the last segment
,
CO
2
– CO
2
concentration to be achieved in the segment,
ramp time
– the time of reaching the set temperature ([d hh:mm]) (days, hours, minutes).
The active value is highlighted in blue.
The item highlighted in red means that the value is out of range and you should enter another one, eg the temperature is above /
below the operating range of the unit or the protection temperature.
The navigation between: segments, program parameters and summary is done by swiping up the segment with the finger or
touch the icon
.
Figure 13 - Program segment edition (1)
Figure 14 - Program segment edition (2)