| Maintenance
64
|
CO2 Incubator Basic
VWR
Chapter 10
Measurement example:
Temperature nominal value: 37 °C
Reference temperature: 36.4 °C
NOTE!
Excessive work space temperature:
Excessive work space temperature after the calibration can be reduced by leaving the doors open
for approx. 30 seconds!
Temperature calibration procedure
1. Activate calibration:
Keep the
key depressed for 5 seconds.
All operating panel indicators flash.
2. Indicate the nominal value:
Press the
key.
The preset value of 37 °C is displayed.
3. Enter the measured value (destination value):
Press the
+
keys.
or
Press the
+
keys.
Destination value e.g. 36.4 °C
4. Accept the destination value:
Press the
key.
The temperature display momentarily shows “CAL”,
Then the corrected actual value (measured destination value 36.4 °C) is displayed.
Содержание CO2 Incubator Basic
Страница 1: ...VWR CO2 Incubator Basic 50150229 Revision 0 December 2015 ...
Страница 54: ... Shut down 52 CO2 Incubator Basic VWR Chapter 8 ...
Страница 62: ... Cleaning and disinfection 60 CO2 Incubator Basic VWR Chapter 9 ...
Страница 74: ... Parts and accessories 72 CO2 Incubator Basic VWR Chapter 11 ...
Страница 77: ...VWR CO2 Incubator Basic 75 Technical data Chapter 12 Average gas consumption Figure 24 CO2 consumption ...
Страница 78: ... Technical data 76 CO2 Incubator Basic VWR Chapter 12 ...
Страница 83: ...VWR CO2 Incubator Basic 81 Certificate of decontamination ...
Страница 84: ... Certificate of decontamination 82 CO2 Incubator Basic VWR Chapter 15 ...
Страница 85: ......
Страница 86: ......