Handling and control |
Chapter 7
Failure code list
Code
Description
Cause
Repair
No values displayed
Communication between
display and CPU- main
board faulty
Contact Service
42
NV RAM read failure
NV RAM faulty,
default values loaded
Contact Service
43
NV RAM read failure
NV RAM faulty,
mirrored values loaded
Contact Service
44
NV RAM faulty
CO
2
measuring cell values
not overwritten, device
works with default values
Contact Service
54
Manipulated variable
failure
Calculation error, device
performed RESET
Contact Service
55
I
2
C bus failure
Data transfer faulty,
measured value quality
dropped below 50%
Eliminate source of interference,
e.g. cellular phone
66
Deviation of temperature
sensors from one another
Temperature signal plau-
sibility doubtful
Contact Service
77
CO
2
cal range exceeded
Max. adjustment value
exceeded
Contact Service
88
Failure upon auto-start
Total time elapsed or max.
adjustment value
exceeded
Repeat auto-start
99
Device doors open
Doors open for more than
10 minutes, door switch
Close device doors, test door
switch for correct function
100
Temperature below
nominal value
Actual value < nominal
value –1 °C
Contact Service
101
Temperature above
nominal value
Actual value > nominal
value +1 °C
Do not exceed ambient
temperature limit
104
Temperature sensor
faulty
Sensor circuit
open/shorted
Contact Service
200
CO
2
below nominal value
Act. val. < nom. val. –1%
No CO
2
Prepressure low
Supply line blocked
Check gas supply:
Connect new gas cylinder
Raise prepressure to 1 bar
Check supply line to device
201
CO
2
above nominal value
Act. val. > nom. val. +1%
Prepressure high
Check gas supply:
Reduce prepressure to 1 bar
204
CO
2
measuring cell
faulty
Sensor circuit
open/shorted
Contact Service
500
Temperature 90°C rou-
tine below nominal value
Actual value < 85°C
Repeat decontamination routine,
contact service as required
501
Temperature 90°C rou-
tine above nominal value
Actual value > 85°C
Contact service
502
Failure in 90°C routine
Power failure during heat-
ing or holding phase
Silence alarm by pressing key
90°C (2 times 5 sec.), then
restart routine
Содержание CO2 Incubator Basic
Страница 1: ...VWR CO2 Incubator Basic 50150229 Revision 0 December 2015 ...
Страница 54: ... Shut down 52 CO2 Incubator Basic VWR Chapter 8 ...
Страница 62: ... Cleaning and disinfection 60 CO2 Incubator Basic VWR Chapter 9 ...
Страница 74: ... Parts and accessories 72 CO2 Incubator Basic VWR Chapter 11 ...
Страница 77: ...VWR CO2 Incubator Basic 75 Technical data Chapter 12 Average gas consumption Figure 24 CO2 consumption ...
Страница 78: ... Technical data 76 CO2 Incubator Basic VWR Chapter 12 ...
Страница 83: ...VWR CO2 Incubator Basic 81 Certificate of decontamination ...
Страница 84: ... Certificate of decontamination 82 CO2 Incubator Basic VWR Chapter 15 ...
Страница 85: ......
Страница 86: ......