background image

14 – VWR STandaRd Line PiPeTTORS 

instruction Manual – 15

Aspirating High-density Liquids 

When pipetting liquids of higher viscosity or lower sur-

face tension than water (e.g., sera or organic solvents), 

a film of liquid may be formed on the inside of the 

pipet tip which may produce erroneous results. as the 

film remains relatively constant in successive pipetting 

operations with the same tip, this error can be elimi-

nated by pre-rinsing the tip and allowing a film to form 

before transferring the first sample. This is achieved by 

aspirating a sample and dispensing it back into the 

same vessel. allowing a film to form prior to sampling 

ensures optimal accuracy and repeatability.
This pre-rinsing operation should be repeated when 

the volume to be aspirated is changed or when a new 

pipet tip is used.

NOTE: 

normally the degree of error resulting from vis-

cous liquids is negligible if pipetting is performed slowly 

and carefully, however, this can be minimized further by 

holding the pipet tip in position for at least 2 seconds 

after aspiration to allow the liquid time to react to the 

change in pressure before it is dispensed. 
if the above method does not result in accurate values, 

recalibrate the pipettor in accordance with Section 9.
it is recommended to record recalibration and correc-

tion values, in order to facilitate reverse calibration to 

a standard liquid.

Filters

The 5,000 and 10,000 μL pipettors include a replaceable 

filter (L), fitted at the bottom of the shaft which prevents 

aspirated liquid from entering the shaft and contami-

nating the inner surface and plunger. Use of the filter is 

particularly important when aspirating and dispensing 

large volumes of liquid. The filter should be replaced if it 

becomes wet.

 9. Checking Pipetting Accuracy Parameters 

   and Pipettor Recalibration

The pipettors have been factory-calibrated using 

gravimetric methods with manufacturer pipet tips and 

distilled water, in accordance with iSO 8655 guidelines 

for the maximum (nominal) liquid volume drawn by the 

pipettor and for 10% of the maximum or minimum liquid 

volume according to the values given (Section 5).
The pipettors are designed to enable recalibration and 

adaptation to different pipetting techniques and liquid 

properties (e.g., temperature, density, and viscosity).

Periodic checks of the operation of the pipettor are 

recommended at least once per year. Frequency of 

checks should be increased depending on workload, 

sterilization or autoclave processes, and frequency of 

replacement of component parts.
if during pipettor operation the accuracy error (the 

difference between the real aspirated volume and the 

preset volume) exceeds the permissible value given 

in the table in Section 5, pipettor recalibration should be 

carried out.
Recalibration of the pipettor involves setting the counter 

to the value of the liquid volume obtained by weighing. 

Recalibration of the pipettor is valid only for one liquid 

volume drawn by the pipettor. 
The recommended volume for recalibration is 10% 

of the maximum (nominal) value or minimum volume 

depending on which of these values is greater. See table 

on page 16 for more details.

Parameters for Checking the Pipetting Accuracy 

The pipetting accuracy is influenced by factors such as: 

tips used, characteristics of the pipetted liquid (density, 

viscosity), and operating conditions (ambient tempera-

ture, pressure).
To determine the accuracy error of the pipettor, the fol-

lowing conditions should be met:

 

ambient temperature and temperature of the pipet-

tor, pipet tips, and liquid should be within the range 

of 20°C to 25°C and stabilized during weighing within 

±0.5°C.

 

Measurements should be conducted using distilled 

water.

 

Balance sensitivity should be suitable for the volume 

“V” to be measured.

Volume Checked (V, μL)

Balance Sensitivity (mg)

0.1≤V≤10

0.001

10≤V≤100

0.01

100≤V≤1,000

0.1

V>1,000

0.1

 

When calculating the liquid volume aspirated by 

the pipettor, the conversion factor (Z) [μL/mg] for dis-

tilled water or a liquid with comparable density should 

be taken into account. Sample values of conversion 

factors are given in the following table. 

Содержание avantor

Страница 1: ...tandard Line Instruction Manual Pipettors VWR Standard Line VWR 8 channel Standard Line VWR 12 channel Standard Line This manual is available in additional languages at www vwr com PMS 647C CMYK 96 54...

Страница 2: ...s are designed for microplate filling The pipet tors enable precise and simultaneous delivery of 8 or 12 preset volumes of liquid These pipettors are available in four volume ranges 0 5 10 L 5 50 L 20...

Страница 3: ...ti channel manifold Contains a set of plungers and a set of flexible suspension shafts which func tion to reduce the force required to attach the tip Model Identification The volume range of the pipet...

Страница 4: ...Accuracy Precision Non filter Tips L Single channel Pipettors 2 Min 0 1 40 0 12 0 10 0 2 12 0 6 0 1 2 7 1 3 Max 2 1 5 0 7 10 Min 0 5 4 0 2 8 10 1 0 2 5 1 8 5 0 1 0 0 6 Max 10 0 0 5 0 4 20 Min 2 3 0 1...

Страница 5: ...o not attach the tips with a rocking movement as this may damage the shaft or plunger Observe this rule particularly with single channel pipettors of low volume range Never draw liquids directly into...

Страница 6: ...In 2 and 10 L pipettors place the cap that can be purchased separately on the bottom of the pipettor shaft and slide the cap upwards to the bottom of the tip ejector M L 10000 L L 10mm 5000 L L 6mm M...

Страница 7: ...ded to pre rinse the tip several times prior to use Do not pipette liquids with temperatures above 70 C NOTE When pipetting acids or corrosive solutions which emit vapors it is recommended to disassem...

Страница 8: ...ISO 8655 guidelines for the maximum nominal liquid volume drawn by the pipettor and for 10 of the maximum or minimum liquid volume according to the values given Section 5 The pipettors are designed t...

Страница 9: ...rm 10 aspirations and calculate the average value in mg Calculate the volume in L by multiplying the value in mg by the conversion factor Z L mg If the average aspirated volume exceeds the permissible...

Страница 10: ...accuracy should not alter if the pipet ting processes including autoclaving are carried out as described in this manual If a change in accuracy occurs it is recommended to Check the calibration of the...

Страница 11: ...ling elements are not sufficiently lubricated e g after repeated autoclaving procedures Incorrect aspiration Liquid with properties other than water den sity viscosity Calibrate the pipettor using the...

Страница 12: ...tional product information and special offers Contact us For information or technical assistance contact your local VWR representative or visit ww vwr com 14 Warranty VWR warrants that this product wi...

Страница 13: ...s Campus Northwest Business Park Ballycoolin Dublin 15 Tel 353 01 88 22 222 Email sales ie vwr com Italy VWR International S r l Via San Giusto 85 20153 Milano MI Tel 39 02 3320311 02 487791 Email inf...

Отзывы: