Instruction Manual INCU-Line Premium Peltier Cooled
Page
42
/
54
You can also select the way of encryption
TLS
or
SSL
.
The above information can be obtained from the email account provider.
Connecting test
Confirms and saves changes
Cancels the entered changes
5.20.
Corrections
In this window
(Figure 47)
you can set correct:
•
The temperature indicated on the display by adding the correction value. The set correction value is
taken in the whole temperature range operation of the device. For example, if the average tempera-
ture displayed by the device indicates 37
°C and the average temperature measured by independent,
external sensor indicates 37
,5°C, the correction should be set on +0,5°C. The average temperature
should be calculated from chosen period of time e.g. 30min. The correction range of -
5°C to +5°C.
Before testing the connection, make sure that the device is connected to a computer net-
work and has a properly configured network connection.
Figure 46 - E-mail
– configuration of the sender e-mail account