7.4.
Access port
A
Ø
30 mm access port can be used to insert an external temperature sensor, which has been secured with
a silicon plug. The plug should cover the access port while the unit is operating. If multiple cables have been
inserted through the access port and if it is not possible to use the plug, secure the access port with adhesive
tape. If you leave the access port open, it may affect temperature stability and uniformity within the chamber.
7.5. Open door alarm
All incubators have been equipped with an open door sensor. If you open the door, icon:
appear on the display.
If you leave the door open for more than 60 seconds, an alarm will sound and the control bar will be red. The
alarm event “door open” is displayed in the list, with the status alarm active.
7.6. Internal LED light
All incubators have been equipped with internal lighting which switches on when opening the door - LED
lights (Philips Corepro LEDspot 3.5-35W GU10 827 36D). Once the light has been switch on, a bulb icon
will appear on the display.
Open door sensor