Instruction Manual INCU-Line Premium
Page
30
/
61
5.9.
Statistics
This panel displays statistics about the currently running program or program that has ended. Statistics are
calculated separately for each segment and separately for each cycle. Data logging starts for calculation
after 30 seconds from reaching the set temperature in the segment. The next data is registered every 1 mi-
nute.
The following information is available:
Set temperature
[°C] – set temperature in the segment
Minimum temperature
[°C] – the lowest recorded temperature
Maximum temperature
[°C] – the highest recorded temperature
Average temperature
[°C] – average temperature
Segment
– status of the segment
-
In progress
– currently performed segment (data is constantly updated)
-
finished
– the segment is made,
-
interrupted
– the segment was interrupted by the User before the set time expired
Segment 1/2
–the number of the currently overviewing segment / number of the currently performed
or completed segment. Navigating between the segments is done by swiping up your finger up or
down.
Cycle 1/1
– the number of currently overviewing cycle / the number of currently performer or com-
pleted segment. Navigating between the segments is done by swiping up your finger up or down.
You cannot overview the segment / cycle data that has not yet started.
Figure 25 - Statistic