www.vtiinstruments.com
EX1629 Command Set
227
vtex1629_self_cal_store
F
UNCTION
P
ROTOTYPE
ViStatus vtex1629_self_cal_store (ViSession
vi
);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
D
ATA
I
TEM
R
ESET
V
ALUE
Not applicable to this function.
D
ESCRIPTION
This function takes the current self-calibration image and stores it to nonvolatile memory, enabling it to be loaded
upon instrument power cycle and reboot. Any other previously stored self-calibration data will be lost. Since the
existence of nonvolatile self-calibration data represents a permanent (although revocable) change from the full
calibration settings, its presence is able to be queried. See the vtex1629_self_cal_is_stored command.
NOTE
Performing a self-calibration does not automatically store the determined calibration constants in
nonvolatile memory. However, it does turn the determined constants into the current self-
calibration data. A call to this function will store that image permanently in non-volatile storage.
E
XAMPLE
ViSession instrumentHandle;
ViStatus status;
…
status = vtex1629_self_cal_store(instrumentHandle);