EX1401 Web Page Operation
available on the network it is connected to, it will instead use Auto IP. Determining the Auto IP
address is discussed in the Network Configuration portion of
Section 2. Other discovery methods,
such as VXI-11, can be used as well.
C
OMMON
I
SSUES
Certain networking issues may prevent instrument discovery, such as connecting the instrument and
host computer to different sub-nets. You may need to contact your network administrator for
assistance with these issues. This section describes possible solutions for other common issues.
Multiple Network Ports
Your computer may have multiple physical or virtual network adaptors. For example, a laptop may
have wi-fi enabled as well as a wired Ethernet connection to the network containing the LXI device.
Some applications also create virtual network adaptors. To resolve this problem, disable all other
network adaptors to force the computer to select the network used by the instrument.
Firewall
A firewall may block some or all network traffic between your computer and the LXI device.
Contact your network administer for assistance with creating exceptions in the firewall
configuration to communication with the instrument. See the table below for a list of ports and
protocols typically used by the EX1401 and other LXI devices.
Protocol /
Service
Port
Base
Protocol
Remarks
mDNS
5353
UDP/TCP
Zero-config protocol for
<hostname>.local address resolution; service discovery;
Multicast 224.0.0251, FF02::FB
HTTP
80
TCP
Instrument web pages (SFP) and driver communication
ICMPv4
ICMP
Typically enables echo request/respond for ping
ICMPv6
ICMP
Optional - Typically enables echo request/respond for ping, also
SLAAC for IPv6 addresses, RDDNS
Arp
Arp
Used to confirm address assignments
DHCPv4
67/68
UDP/TCP
IPv4 address assignment, DNS Server, Dynamic DNS,
Gateway
DNS
53
UDP
Naming service
SSH
22
TCP
Bidirectional interactive text oriented communication
RPC port-
mapper
111
UDP/TCP
Builds upon Sun-RPC and port-mapper
VXI-11
Varies
TCP
VXI-11 instrument discovery
The exact port number varies; query the RPC port-mapper to
determine which port is current in use
LXI-eventsvc
5044
UDP/TCP
LXI Event support for instrument triggering; multicast
224.0.23.159, FF02::138. LXI Events can use other ports also,
but default port number is 5044
Ptp-event
319
UDP/TCP
LXI Profile IEEE 1588
Precision Time Protocol (PTP);
multicast 224.0.1.129, FF02::181
Ptp-general
320
UDP/TCP
LXI Profile IEEE 1588 Precision Time Protocol (PTP);
multicast 224.0.1.129, FF02::181
Data stream
9900
TCP
Driver data stream
T
ABLE
4-1:
S
TANDARD
P
ORTS
,
P
ROTOCOLS
,
AND
S
ERVICES
G
ENERAL
W
EB
P
AGE
O
PERATION
When initial connection is made to the EX1401, the instrument home page, or
Index
, appears. This
page displays instrument-specific information. This page is accessible from any other instrument
page by clicking on the EX1401 web page header. The EX1401 Navigation Menu is displayed on
Содержание EX1401
Страница 2: ......
Страница 24: ......
Страница 57: ...www vtiinstruments com EX1401 Web Page Operation 57 ...
Страница 94: ......
Страница 98: ......
Страница 102: ......
Страница 111: ...www vtiinstruments com EX10xxA Index 111 FIGURE A 1 IEEE1588 NETWORK TOPOLOGY ...
Страница 112: ......
Страница 114: ...VTI Instruments Corp 114 EX1401 Index ...