www.vtiinstruments.com
EX10xxA Function Calls
267
vtex10xxA_set_vtb_wiredor
F
UNCTION
P
ROTOTYPE
ViStatus vtex10xxA_set_vtb_wiredor (ViSession
vi
,ViInt32
wiredor
);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
wiredor
= an integer input value that indicates the Wired-OR state for all eight LXI Trigger Bus lines. Only the
least significant bits are used. The least significant bit of the integer represents LXI Trigger Bus line 0, where the 8th
least significant bit corresponds to LXI Trigger Bus line 7. A return value of “1” for any of the eight bits indicates
that the wired or mode is enabled for that particular LXI Trigger Bus line, while a value of “0” indicates that Wired-
OR is disabled.
D
ATA
I
TEM
R
ESET
V
ALUE
wiredor
= 0
D
ESCRIPTION
This function sets the Wired-OR state for each of the individual LXI Trigger Bus lines.
E
XAMPLE