VTI Instruments Corp.
208
EX10xxA/RX1032/EX1044 Function Calls
vtex10xxA_reset
F
UNCTION
P
ROTOTYPE
ViStatus vtex10xxA_reset(ViSession
vi);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
D
ATA
I
TEM
R
ESET
V
ALUE
Not applicable to this function.
D
ESCRIPTION
This function performs an instrument reset, returning all of the EX10xxA’s acquisition configuration parameters to
their default values.
NOTE
An instrument reset clears the FIFO reading memory. All desired acquisition data must be retrieved
from the FIFO prior to the issuance of this function.
E
XAMPLE
Содержание EX1000A