Calibration Procedure
1)
Connect to the embedded web page. See
2)
Click
Soft Front Panel
from the
Index
page to navigate to the
Soft Front Panel
page. Select
the
EX1200-dmm DMM
from the list of available plug-in modules (the slot designation will
vary depending on the mainframe being used).
F
IGURE
9-6:
I
NSTRUMENT
SFP
S
ELECTION
P
AGE
3)
At the top of the
DMM’s
soft front panel, click the
Calibration
button.
F
IGURE
9-7:
C
ALIBRATION
T
AB
4)
From the
Calibration Functions
section of the
Calibration
page, select
Calibration
from
the drop down menu.
F
IGURE
9-8:
C
ALIBRATION
F
UNCTION
S
ELECTION
Select the DMM
Calibration Tab
Содержание Ametek EX1200A Series
Страница 17: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 18: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 33: ......
Страница 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Страница 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Страница 65: ......
Страница 73: ......
Страница 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Страница 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Страница 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Страница 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Страница 155: ......
Страница 161: ......