Appendix
82
EzRay Air Portable (Model: VEX-P300) User Manual
A.3
Tables of Exposure Times (Default)
The following exposure timetables were established with a unit equipped with a cone
that corresponds to a focus-to-skin distance of 200 mm (8 inches), respectively.
DIGITAL SENSOR
1) Option 1: 60 kV, 2.5 mA, 0.05~1.0 sec
Patient
Teeth
Angle of
inclination
SSD: 200 mm (8 inches)
kV
mA
s
Adult
Incisor
Maxilla: +45˚
Mandible: -
25˚
60
2.5
0.14
Canine
Maxilla: +45˚
Mandible: -
20˚
60
2.5
0.16
Molar/
Premolar
Maxilla: +30˚
Mandible: -
5˚
60
2.5
0.18
Bitewing
+5˚~ +8˚
60
2.5
0.20
Child
Incisor
Maxilla: +45˚
Mandible: -
25˚
60
2.5
0.12
Canine
Maxilla: +45˚
Mandible: -
20˚
60
2.5
0.14
Molar/
Premolar
Maxilla: +30˚
Mandible: -
5˚
60
2.5
0.16
Bitewing
+5˚~ +8˚
60
2.5
0.18
Содержание EzRay Air Portable
Страница 2: ......
Страница 4: ...Notice ii EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...
Страница 24: ...3 System Overview 22 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...
Страница 48: ...4 Operation 46 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...
Страница 64: ...5 Service Mode 62 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...
Страница 72: ...8 Disposing of the Unit 70 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...
Страница 76: ...9 Device Specifications 74 EzRay Air Portable Model VEX P300 User Manual 4 Tube Dimensions mm ...
Страница 101: ...Appendix EzRay Air Portable Model VEX P300 User Manual 99 ENGLISH This page is intentionally left blank ...
Страница 102: ...Appendix 100 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...