2.9 Operation
Simulator
Only switch on the simulator if the simulator temperature and the room temperature
are between 10–30 °C (50–86 °F).
The power inlet of the simulator is located at the base of the lifting column near the
floor. Make sure that water never comes into contact with the power inlet. Advise
cleaning personnel to be especially careful while wet mopping the floor near the
simulator. It has to be made sure that the floor is not left even slightly wet near the
power inlet. Otherwise electric shock hazard or device damage may occur.
Remove the power plug immediately if liquid enters any part of the enclosure. In
this case, the equipment needs to be checked by VRmagic or an authorized service
technician.
When opening the housing of any of the simulator components while the system
is connected to power, there is the risk of death or serious injury through electrical
shock. Further, the simulator might get damaged and the warranty will be voided.
Never attempt to open the housing of any of the components.
In the event of failure, or if safe operation of the simulator or any device connected
to it cannot be assured, the simulator must be taken out of operation and must be
secured against unintentional activation or use. Disconnect the power plug from
the power source immediately. Ask VRmagic or an authorized service technician to
check the device.
Safe operation is not assured
• if the voltage requirements of the simulator do not match the power supply
voltage.
• if any part of the simulator or a cable is visibly damaged.
• if the simulator is not operating properly.
• if the simulator has been exposed to rain or other wet conditions.
• if liquid or objects have entered any part of the simulator.
• after storage under adverse conditions.
• after heavy strain in transporting the device.
The slit lamp uses infrared LEDs which are invisible to the human eye. The LEDs
are positioned in the lower rear part of the slit lamp base and are only active while a
case is running. Looking directly into the LEDs may cause irritation or damage to the
eye. Do not look directly into the infrared LEDs from a short distance for a long time.
If the product is used as intended and described in the product documentation, the
infrared LEDs present no danger.
The simulator uses components that are sensitive to magnetism. If magnets are
brought too close to the simulator, the components may get damaged permanently
or temporary errors may occur during the simulation. Always keep magnets away
from the slit lamp.
11
Safety
Eyesi Slit Lamp Simulator – Installation and Operating Guide
Содержание Eyesi Slit Lamp
Страница 1: ...Eyesi Slit Lamp Installation and Operating Guide Platform 1 0 ...
Страница 3: ......
Страница 60: ...60 Notes Eyesi Slit Lamp Simulator Installation and Operating Guide ...
Страница 61: ...61 Notes Eyesi Slit Lamp Simulator Installation and Operating Guide ...
Страница 62: ...62 Notes Eyesi Slit Lamp Simulator Installation and Operating Guide ...
Страница 63: ......