C H A P T E R 1 4 - A T E D I T O R
AT5600 User Manual 98-119 issue 14
Page 365
14.6.37.
LSBX
–
Inductance with External Bias (Series
Circuit)
DC Bias
Measure
Range
Test
Voltage
Test
Frequency
Basic Accuracy
0.1A
to 1000A
1nH 1MH
1mV to 5V
20Hz to 3MHz
0.2% @ Q>10
This test requires the use of one or more Voltech DC1000 Precision DC Bias
Current Sources. Contact your Voltech sales office for details.
The inductance of a transformer winding while an external bias current is flowing
through it may be tested using series or parallel equivalent circuit models. Initially
the DC bias current is set up and allowed to stabilize. An AC voltage is applied
across the selected winding; the voltage across and current through the winding is
then measured using harmonic analysis. The measured voltage is divided by the
current to obtain a complex impedance and the inductance is calculated.
On selecting ‘LSBX Inductance with External Bias (Series Circuit)’ from the
‘Available Tests’ window, a dialogue box will be displayed.
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...