C H A P T E R 1 0
–
S P E C I F I C A T I O N S
AT5600 User Manual 98-119 issue 14
Page 225
10.2.17.
SURG Test
The measurement accuracy is based on the chosen test conditions, and is given
by:
𝐴
𝑇
= 𝐴
𝑅
+ 𝐴
𝐶
+ 𝐴
𝑉
where:
A
T
is the total accuracy (%)
A
R
is the basic relative accuracy (%)
A
C
is the calibration accuracy (%)
A
V
is the correction for voltage level (%)
where:
𝐴
𝑅
= 3.00 %
𝐴
𝐶
= 3.00 %
𝐴
𝑉
= 2.0 % + (
250 %
𝑉
𝑃
) + (0.001 % ∗ 𝑉
𝑃
)
where:
V
P
is the programmed peak voltage (V)
Maximum programmable voltage
For UUTs with Ls => 125uH, the unit maximum is achievable, namely 5000V DC
For UUTs with Ls<=125uH, the maximum voltage achievable is defined by
Vout = (7000 * Luut) / (Luut + 50)
Where Luut = UUT Series Inductance in uH
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...