C H A P T E R 1 0
–
S P E C I F I C A T I O N S
AT5600 User Manual 98-119 issue 14
Page 223
10.2.15.
WATT and STRW Tests
The accuracy of the applied test voltage is 2%.The measurement accuracy is
based on the chosen test conditions, and is given by:
𝐴
𝑇
= 𝐴
𝑅
+ 𝐴
𝐶
+ 𝐴
𝑉
+ 𝐴
𝐼
+ 𝐴
𝑃𝐹
where:
A
T
is the total accuracy (%)
A
R
is the basic relative accuracy (%)
A
C
is the calibration accuracy (%)
A
V
is the correction for voltage level (%)
A
I
is the correction for current level (%)
A
PF
is the correction for power factor (%)
where:
𝐴
𝑅
= 0.30 % 𝑓𝑜𝑟 𝑊𝐴𝑇𝑇 𝑜𝑟 1.00 % 𝑓𝑜𝑟 𝑆𝑇𝑅𝑊
𝐴
𝐶
= 0.80 % + (
0.001 %
𝐹𝑚
)
𝐴
𝑉
= 0.05 % + (
0.01 %
𝑉
𝑀
)
𝐴
𝐼
= 0.05 % + (
0.01 %
𝐼
𝑀
) + (1.0 % ∗ 𝐼𝑚)
𝐴
𝑃𝐹
= 1.0 % ∗ (
√(1 − 𝑃𝐹
2
)
𝑃𝐹
)
where:
F
M
is the test frequency (Hz)
V
M
is the measured voltage (V)
I
M
is the measured current (A)
PF
is the power factor of the load
and:
𝑃𝐹 =
𝑊
𝑉𝐴
where:
W
is the measured power (W)
VA
is the product of test voltage and magnetizing current
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...