C H A P T E R 1 0
–
S P E C I F I C A T I O N S
AT5600 User Manual 98-119 issue 14
Page 209
10.2.
Accuracy Specifications – Available Tests
10.2.1.
R Test
The accuracy is based on the chosen test conditions, and is given by:
𝐴
𝑇
= 𝐴
𝑅
+ 𝐴
𝐶
+ 𝐴
𝑉
+ 𝐴
𝐼
where:
A
T
is the total accuracy (%)
A
R
is the basic relative accuracy (%)
A
C
is the calibration accuracy (%)
A
V
is the correction for voltage level (%)
A
I
is the correction for current level (%)
where:
𝐴
𝑅
= 0.10%
𝐴
𝐶
= 0.08%
𝐴
𝑉
= 0.03% +
0.1%
𝑉
𝑀
𝐴
𝐼
= 0.03% +
0.001%
𝐼
𝑀
where:
V
M
is the measured voltage (V)
I
M
is the measured current (A)
The test signal is set according to the value specified as maximum in the test
limits: -
Where:
V = Volts (V)
I = Current (A)
R = Resistance (
Ω
)
Maximum
Resistance R
<1
Ω
1
Ω
-10k
Ω
10k
Ω
-
50k
Ω
>50k
Ω
Test current
I=1
I=1/R
V=R*100u
V=5
Содержание AT5600
Страница 10: ...AT5600 User Manual 98 119 issue 14 Page 10 16 4 AT Series Server Software 437...
Страница 39: ...C H A P T E R 2 S Y S T E M O V E R V I E W AT5600 User Manual 98 119 issue 14 Page 39...
Страница 88: ...C H A P T E R 6 S A F E T Y S Y S T E M S AT5600 User Manual 98 119 issue 14 Page 88...
Страница 168: ...C H A P T E R 7 T E S T S A N D T E S T C O N D I T I O N S AT5600 User Manual 98 119 issue 14 Page 168...
Страница 198: ...C H A P T E R 8 F R O N T P A N E L O P E R A T I O N AT5600 User Manual 98 119 issue 14 Page 198...
Страница 204: ...C H A P T E R 9 T R O U B L E S H O O T I N G AT5600 User Manual 98 119 issue 14 Page 204...
Страница 252: ...C H A P T E R 1 0 S P E C I F I C A T I O N S AT5600 User Manual 98 119 issue 14 Page 252...
Страница 258: ...C H A P T E R 1 2 W A R R A N T Y A N D S E R V I C E AT5600 User Manual 98 119 issue 14 Page 258...
Страница 270: ...C H A P T E R 1 3 T E S T F I X T U R E S AT5600 User Manual 98 119 issue 14 Page 270...