93
Alternating current (A/AC)
Range
Resolution
Accuracy
600.00 µA
0.01 µA
45 Hz - 1 kHz ±(0.5% + 30)
>1 kHz - 10 kHz ±(1.0% + 30)
6000.0 µA
0.1 µA
45 Hz - 1 kHz ±(0.5% + 30)
>1 kHz - 10 kHz ±(1.0% + 30)
60.000 mA
0.001 mA
45 Hz - 1 kHz ±(0.5% + 30)
>1 kHz - 10 kHz ±(1.0% + 30)
600.00 mA
0.01 mA
45 Hz - 1 kHz ±(0.5% + 30)
>1 kHz - 10 kHz ±(1.0% + 30)
6.0000 A
0.0001A
45 Hz - 1 kHz ±(0.5% + 30)
>1 kHz - 10 kHz ±(1.0% + 30)
10.000 A
0.001 A
45 Hz - 1 kHz ±(0.5% + 30)
>1 kHz - 10 kHz ±(1.0% + 30)
Overload protection: Fuse
Specified measurement range: 10–100 % of the measurement range
Fuses: µA/mA = F600mAH1000V high-performance ceramic fuse
10 A = F10AH1000V high-performance ceramic fuse
Measuring time 10 A input: 10 seconds with 15-minute intervals
TrueRMS peak (Crest Factor (CF)) ≤3 CF over the entire range
TrueRMS peak for non-sinusoidal signals plus tolerance:
CF >1.0 - 2.0 + 3%
CF >2.0 - 2.5 + 5%
CF >2.5 - 3.0 + 7%