87
17 Technical data
Display ...........................................................4000 Counts (digits)
Measurement interval ....................................approx. 2-3 measurements/second
AC measurement method ..............................True RMS, AC-coupled
Test lead length .............................................approx. 90 cm
Measuring impedance ...................................
≥10 MΩ (400 mV: ≥100 MΩ)
Measuring terminal clearance .......................19 mm (COM-V)
Low battery indicator .....................................Battery voltage <3.6 ±0.2 V
“Dangerous voltage” indicator .......................
≥30 V/AC-DC
“Range exceeded” alarm ...............................
≥600 V/AC-DC, ≥10 A/AC-DC
“OL” (overload) alarm ....................................
≥610 V/AC-DC, ≥10,10 A/AC-DC or
measurement >4000 counts
Automatic power-off ...................................... after approx.15 minutes (can be
manually disabled)
Current consumption (auto off) ......................<50 µA
Operating voltage ..........................................3 x AAA 1.5 V batteries
Operating conditions ......................................0 to +40 °C (<75 % RH)
Operating altitude ..........................................max. 2000 m above sea level
Storage temperature ......................................-10 °C to +50 °C
Weight ...........................................................approx. 375 g
Dimensions (L x W x H) .................................190 x 90 x 43 mm
Measuring category .......................................CAT III 600 V
Pollution degree .............................................2
Operating environment ..................................Indoor use
Safety regulations ..........................................EN 61010-1 and EN61010-2-033
F1 FUSE ........................................................
Φ6.35×32 mm, FF 10 A, H 600 V,
breaking capacity: 10 kA
F2 FUSE ........................................................
Φ5×20 mm, FF 2.5 A, H 700V, break
-
ing capacity: min. 300 A