22
IN OFFICE PROGRAMMING
The SenTiva® (M1000) VNS Therapy System offers a number of new features
that help simplify and customize patient care.
The generator’s output current for the programmed stimulation in all modes
must be 0 mA for the first 14 days after implantation.
Manual Programming
Manually adjust parameters
Prone Position/Low Heart Rate Detection
Monitor the patient for episodes of low heart rate and prone body position
after an AutoStim or magnet activation
Guided Programming
Simplifies programming with option of Standard or Custom
Therapy Protocols
Scheduled Programming
Automatically change parameter steps according to a preset schedule
Day/Night Programming
Delivers two independent sets of therapy parameter schedules at different
times during a 24-hour period
NOTE:
The programmer must be IN SESSION to access these features.
NOTE:
At the end of each office visit, the generator must be interrogated by
tapping the
INTERROGATE
button followed by tapping the
END SESSION
button to ensure creation and accuracy of the session reports.