Alarms
This chapter describes the alarm functions used for the Vivo 65.
Permitted distributed alarm systems are Vivo 65 remote alarm with cable
and Vivo 65 nurse call cables provided by Breas Medical AB only.
6.1 Alarm Function
The alarm function of the Vivo 65 consists of the alarm LEDs on the
front panel, an audible alarm, and messages on the display. (See “The
Vivo 65's Front Panel” on page 23 for an overview of the position of the
LEDs.)
Alarm Indication
When an alarm condition arises, the alarm is indicated in three ways:
The adjustable alarm settings should be re-evaluated whenever a change
in settings is made on the Vivo 65.
• Never leave a patient unattended during an alarm condition.
• Setting alarm limits to extreme values could put the patient at risk.
The alarm settings are maintained during an extended power failure.
As soon as an alarm condition is detected, the Vivo 65 main unit and (if
connected) the remote alarm unit will alarm without delay.
• Color LED on the panel: indi-
cates the priority of the active
alarm condition.
– High priority: red color, flash-
ing twice per second.
– Medium priority: yellow color,
flashing every 2 seconds.
Содержание 65
Страница 202: ...202 Accessories Vivo 65 users manual Doc 006088 En US U 1b 15 14 19 21 22 24 25 18 20 27 13 17 16 26 28 29 12 23 ...
Страница 208: ...208 Patient Settings Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 210: ...210 FAA Compliance Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 211: ...211 FAA Compliance Vivo 65 users manual Doc 006088 En US U 1b ...
Страница 228: ...228 Index Vivo 65 users manual Doc 006088 En US U 1b W Warning icon 6 Water trap 17 Weight Vivo 191 ...